metrology.calibration

Calibration utilities for mapping digital signals to physical SPL levels.

Auto-generated from the source docstrings by scripts/generate_api_docs.py (make api-docs). Do not edit by hand.

calculate_sensitivity(
ref_signal: list[float] | np.ndarray,
target_spl: float = 94.0,
ref_pressure: float = 2e-05,
fs: int | None = None,
validate: bool = True,
max_fluctuation_db: float | None = None,
frequency: float = 1000.0,
narrowband: bool = False,
) -> float

Deprecated alias of sensitivity.

The calibration reference recording looks unreliable.

sensitivity(
ref_signal: list[float] | np.ndarray,
target_spl: float = 94.0,
ref_pressure: float = 2e-05,
fs: int | None = None,
validate: bool = True,
max_fluctuation_db: float | None = None,
frequency: float = 1000.0,
narrowband: bool = False,
) -> float

Calculate the calibration factor (multiplier) to convert digital units to Pascals based on a reference recording (e.g., 1kHz @ 94dB).

When fs is provided (and validate is True), the recording’s stability is checked the way IEC 60942:2017 specifies for the calibrator itself (5.3.3): levels are measured with time-weighting F and the short-term level fluctuation (the absolute difference between each of the maximum and minimum levels and the mean level) must not exceed the Table 2 acceptance limit for the calibrator class (class 1: 0.07 dB at and above 160 Hz, relaxed to 0.10 dB above 63 Hz and below 160 Hz, and to 0.20 dB for the 31,5-63 Hz rows where the F time-weighting itself ripples; below Table 2’s 31,5 Hz span the strict 0.07 dB applies). A larger fluctuation usually means a badly coupled microphone or handling noise in the recording, which would silently corrupt every calibrated level; a CalibrationWarning is emitted.

Parameters

NameDescription
ref_signalRecording of the calibration tone.
target_splThe known SPL level of the calibrator (default 94 dB).
ref_pressureReference pressure (default 20 microPascals).
fsSample rate of the recording in Hz. Required for the stability validation; without it the check is skipped.
validateIf True (default) and fs is given, warn when the recording’s short-term level fluctuation exceeds the limit.
max_fluctuation_dbExplicit fluctuation limit in dB. Default (None) resolves the IEC 60942:2017 Table 2 class 1 limit for frequency.
frequencyNominal frequency of the calibration tone in Hz (default 1000.0), used to select the Table 2 row.
narrowbandIf True (requires fs), estimate the tone level with a coherent single-frequency (Goertzel) detector locked to the tone near frequency instead of the full-band RMS. This rejects broadband hum/noise in the reference take, which otherwise inflates the RMS and shrinks the factor by -10*lg(1 + 1/SNR) (about -0.44 dB at 20 dB SNR), silently biasing every subsequent level. The default (False) keeps the exact legacy broadband-RMS behaviour; enable it for noisy coupler recordings.

Returns: Calibration factor (sensitivity multiplier).