Calibration utilities for mapping digital signals to physical SPL levels.
Auto-generated from the source docstrings by
scripts/generate_api_docs.py(make api-docs). Do not edit by hand.
calculate_sensitivity
Section titled “calculate_sensitivity”calculate_sensitivity( ref_signal: list[float] | np.ndarray, target_spl: float = 94.0, ref_pressure: float = 2e-05, fs: int | None = None, validate: bool = True, max_fluctuation_db: float | None = None, frequency: float = 1000.0, narrowband: bool = False,) -> floatDeprecated alias of sensitivity.
CalibrationWarning
Section titled “CalibrationWarning”The calibration reference recording looks unreliable.
sensitivity
Section titled “sensitivity”sensitivity( ref_signal: list[float] | np.ndarray, target_spl: float = 94.0, ref_pressure: float = 2e-05, fs: int | None = None, validate: bool = True, max_fluctuation_db: float | None = None, frequency: float = 1000.0, narrowband: bool = False,) -> floatCalculate the calibration factor (multiplier) to convert digital units to Pascals based on a reference recording (e.g., 1kHz @ 94dB).
When fs is provided (and validate is True), the recording’s
stability is checked the way IEC 60942:2017 specifies for the calibrator
itself (5.3.3): levels are measured with time-weighting F and the
short-term level fluctuation (the absolute difference between each of
the maximum and minimum levels and the mean level) must not exceed the
Table 2 acceptance limit for the calibrator class (class 1: 0.07 dB at
and above 160 Hz, relaxed to 0.10 dB above 63 Hz and below 160 Hz, and to
0.20 dB for the 31,5-63 Hz rows where the F time-weighting itself ripples;
below Table 2’s 31,5 Hz span the strict 0.07 dB applies). A larger
fluctuation usually means a badly coupled microphone or handling noise in
the recording, which would silently corrupt every calibrated level; a
CalibrationWarning is emitted.
Parameters
| Name | Description |
|---|---|
ref_signal | Recording of the calibration tone. |
target_spl | The known SPL level of the calibrator (default 94 dB). |
ref_pressure | Reference pressure (default 20 microPascals). |
fs | Sample rate of the recording in Hz. Required for the stability validation; without it the check is skipped. |
validate | If True (default) and fs is given, warn when the recording’s short-term level fluctuation exceeds the limit. |
max_fluctuation_db | Explicit fluctuation limit in dB. Default (None) resolves the IEC 60942:2017 Table 2 class 1 limit for frequency. |
frequency | Nominal frequency of the calibration tone in Hz (default 1000.0), used to select the Table 2 row. |
narrowband | If True (requires fs), estimate the tone level with a coherent single-frequency (Goertzel) detector locked to the tone near frequency instead of the full-band RMS. This rejects broadband hum/noise in the reference take, which otherwise inflates the RMS and shrinks the factor by -10*lg(1 + 1/SNR) (about -0.44 dB at 20 dB SNR), silently biasing every subsequent level. The default (False) keeps the exact legacy broadband-RMS behaviour; enable it for noisy coupler recordings. |
Returns: Calibration factor (sensitivity multiplier).