vibration.human.instrumentation

Type-testing a human-vibration meter against ISO 8041-1:2017.

A sound level meter can be given a class: IEC 61672-1 prints design goals and tolerance limits, and verify_weighting_class turns a measured response into a verdict against them. A human-vibration meter is checked the same way and by the same kind of table, and this module is that check.

What the standard grades. ISO 8041-1 defines the nine frequency weightings (WEIGHTING_NAMES) as exact transfer functions, and then says how far a real instrument may sit from them. The allowance is not one number: it is a band that widens away from the middle of the working range, keyed to four transition frequencies per weighting (Table 4). Table 5 prints five rows across them and they carry three distinct limit pairs, because the two skirts share theirs and so do the two tails: inside the central region the magnitude may differ by +12 % / −11 %; in the two skirts by +26 % / −21 %; beyond the outermost pair the standard stops constraining the response from below altogether, which is what −100 % means in the printed table.

What it does not grade. A verdict here is about the frequency weighting, which is one clause of a standard that also covers indication, linearity, overload, temperature, humidity and electromagnetic susceptibility. Passing this check is a necessary condition for conformity, never a certificate of it: the rest are laboratory tests on hardware, not arithmetic.

Tolerances are on the factor, not on the decibel. The standard writes its limits as percentages of the weighting factor, and this module keeps them that way. +26 % is +2,0 dB to two decimals, but the percentage is what the page prints and what the acceptance test in Annex B is written in.

The band does not widen; the measurement does. Two sentences of the standard talk about expanded uncertainty and they are not the same sentence. 5.6.6 (printed folio 14) says the Table 5 limits already “include the applicable maximum expanded uncertainties of measurement”, which is why weighting_tolerance_percent returns the printed numbers and never adds to them. 13.1 (folio 42) and 14.1 (folio 48) say something else, word for word in both: compliance is demonstrated when the measured deviation, “extended by the actual expanded uncertainty of measurement of the testing laboratory”, does not exceed those limits. That second rule is about the laboratory’s own uncertainty, it moves the deviation rather than the band, and it is what expanded_uncertainty_percent does in verify_weighting.

Phase is graded on a slope, not on an angle. Footnote a of Table 5 limits the phase criterion to instruments reporting a parameter not based on r.m.s. values, and 5.6.6 explains why the criterion is not the phase error itself: a constant group delay is a large phase error that changes no measured quantity. Formula (6) (folio 15) turns a pair of adjacent phase errors into the characteristic phase deviation the table actually grades, and verify_phase_response is that comparison.

The personal meter of Part 2. ISO 8041-2:2021 specifies the personal vibration exposure meter (PVEM), the instrument left unattended to log a worker’s exposure through a full working day, with its transducer on the worker, the seat or the machine (5.1.1 to 5.1.3, folios 4 and 6), and it grades that instrument with the tables of Part 1 wherever it can: the weightings of Table 3, the bands of Tables 4 and 5 and the saw-tooth burst of Tables 6 to 9 are printed again unchanged, which is why every function here serves both parts. Two things differ, and each has a table of its own. Its Table 2 (folio 8) keeps the first two rows of the Part 1 table and drops the third, because its 5.13 (folio 15) declares the running r.m.s. “Not applicable for PVEM”: PVEM_INDICATION_TOLERANCES_PERCENT. And its clauses 12 and 13 print their own maximum expanded uncertainties, the same figures as Part 1 on a shorter list of clauses: PVEM_MAX_EXPANDED_UNCERTAINTY_PERCENT.

Auto-generated from the source docstrings by scripts/generate_api_docs.py (make api-docs). Do not edit by hand.

band_limited_weighting_factor(name: str) -> float

The factor that makes the second row of Table 2 satisfiable.

Table 2 (folio 12) allows 3 % between “the indicated value of any frequency-weighted measurement quantity” and “the indicated value of the corresponding band-limiting measurement multiplied by the appropriate weighting factor”, and 12.7 (folio 30) turns it into a procedure: with the input adjusted so that the meter indicates the reference vibration value with band-limiting frequency weighting, the frequency-weighted indication “shall equal the indicated band-limited weighted vibration value multiplied by the appropriate weighting factor (see Table 1)”.

That test fixes the input at a_ref / |H_BL(f_ref)|, so the weighted indication is a_ref |H(f_ref)| / |H_BL(f_ref)|: the factor that makes the identity true is the ratio of the two responses at the reference frequency, which is what this function returns.

========== ================== =================== Weighting This ratio Table 1 prints ========== ================== =================== Wb 0,812 819 0,812 6 Wc 0,514 617 0,514 5 Wd 0,126 120 0,126 1 We 0,062 891 0,062 87 Wf 0,418 982 0,388 8 Wh 0,202 025 0,202 0 Wj 1,018 841 1,019 Wk 0,772 066 0,771 8 Wm 0,336 336 0,336 2 ========== ================== ===================

For eight of the nine the distinction is academic: their band-limiting weighting is between 0,999 68 and 0,999 97 at their reference frequency, so the printed Table 1 factor is the same number to 0,03 %, well inside any of the Table 2 limits the row is judged against. For Wf it is not. Its reference frequency, 2,5 rad/s = 0,397 887 Hz, sits inside its own band-limiting skirt (0,08 Hz and 0,63 Hz corners, Table 3): the band-limiting weighting is 0,928 078 there and the overall weighting 0,388 848, values Table B.5 prints as 0,927 9 and 0,388 4 at the neighbouring 0,398 1 Hz band centre. Reading “the appropriate weighting factor” as the 0,388 8 of Table 1 makes a conforming Wf meter miss the row by 7,76 % of the value it demands, half again over the ±5 % Table 2 allows it; reading it as the ratio 0,418 982 makes the row true by construction. The standard does not define the phrase, and the “(see Table 1)” of 12.7 points at the reading that cannot be satisfied; the ambiguity is registered in docs/ERRATA.md.

Parameters

NameDescription
nameOne of WEIGHTING_NAMES.

Returns: |H(f_ref)| / |H_BL(f_ref)|, dimensionless.

Raises

ExceptionWhen
ValueErrorIf the weighting is not one of the nine.

Constant (tuple).

CENTRAL_TOLERANCE_PERCENT = (12.0, -11.0, 6.0)
characteristic_phase_deviation(
frequencies_hz: ArrayLike,
phase_deviation_deg: ArrayLike,
) -> NDArray[np.float64]

The characteristic phase deviation of ISO 8041-1 Formula (6).

5.6.6 (printed folio 14) says why the phase error itself is not the criterion: “the errors in measurement due to errors in the phase response are dependent on the rate of change in phase error with frequency, rather than the absolute phase error itself”. Formula (6) (folio 15) is that rate, printed inside absolute-value bars:

The normative Annex H closes the two questions the clause leaves open. Formula (H.3) (folio 93) prints the same quantity with the two products exchanged, which is the same number inside the bars both formulae carry, and it says where the number belongs: “This allows the calculation of Δφ0(f_n) at each frequency f_n except for the highest frequency”. So N frequencies give N - 1 values, each attributed to the lower frequency of its pair, and that is what decides which Table 5 region grades a pair that straddles a transition frequency: the region of f_n.

Two closed forms say what the quantity measures. A constant phase error of c degrees gives |c| at every pair, so an offset is graded at face value. A phase error proportional to frequency, which is a constant group delay, gives exactly zero, and NOTE 1 of H.2.1 is that reading: a constant group delay “would probably far exceed the tolerances on phase deviation, but would influence neither the vibration parameters to be measured nor the characteristic phase deviation values”.

Parameters

NameDescription
frequencies_hzThe frequencies the phase errors belong to, in hertz, strictly ascending. H.2.1 asks for them “preferably in steps of one-third octaves”, which is the grid Annex B tabulates.
phase_deviation_degThe phase error at each frequency, in degrees, measured minus design goal.

Returns: One value per adjacent pair, in degrees, attributed to the lower frequency of the pair, so of length one less than the input.

Raises

ExceptionWhen
ValueErrorIf the two arrays do not have the same shape, if a frequency is not positive and finite, if a phase deviation is not finite, or if there are fewer than two frequencies or they do not strictly ascend.

Constant (float).

INDICATION_TOLERANCE_PERCENT = 4.0
indication_tolerance_percent(name: str) -> float

The Table 2 indication tolerance at the reference frequency.

Parameters

NameDescription
nameOne of WEIGHTING_NAMES.

Returns: The permitted deviation of the indication, in per cent.

Raises

ExceptionWhen
ValueErrorIf the weighting is not one of the nine.

Constant (float).

ISO8041_COVERAGE_FACTOR = 2.0

LOW_FREQUENCY_INDICATION_TOLERANCE_PERCENT

Section titled “LOW_FREQUENCY_INDICATION_TOLERANCE_PERCENT”

Constant (float).

LOW_FREQUENCY_INDICATION_TOLERANCE_PERCENT = 5.0

Constant (str).

LOW_FREQUENCY_WEIGHTING = 'Wf'

Constant (mapping).

MAX_EXPANDED_UNCERTAINTY_PERCENT = {'12.7': 2.0, '12.10.1': 2.0, '12.10.2': 3.0, '12.10.2 additional ranges': 4.0, '12.11.2': 4.5, '12.11.3': 3.0, '12.11.4': 5.0, '12.13': 3.0, '12.14': 2.0, '12.18': 0.01, '13.9': 2.0, '13.11': 4.0, '13.14': 2.0, '13.15': 0.01, '14.9': 5.0}

Constant (mapping).

NOMINAL_FREQUENCY_RANGE_HZ = {'Wb': (0.5, 80.0), 'Wc': (0.5, 80.0), 'Wd': (0.5, 80.0), 'We': (0.5, 80.0), 'Wf': (0.1, 0.5), 'Wh': (8.0, 1000.0), 'Wj': (0.5, 80.0), 'Wk': (0.5, 80.0), 'Wm': (1.0, 80.0)}
peak_deviation_percent(
characteristic_phase_deviation_deg: ArrayLike,
) -> float

The peak-value deviation a phase response costs (Formula (H.4)).

Annex H is normative, and (H.4) (folio 93) is the only worked number in the whole phase argument:

followed by “For the maximum characteristic phase deviations of 12°, the maximum peak value deviation is approximately 10 %”. The maximum is inside the printed formula, so this returns one number for a whole response rather than one per frequency: ΔP_max is what the worst pair costs.

NOTE 2 of H.2.1 fences the approximation twice, and both fences matter to a reader of the returned number. It “applies to small Δφ0 values only (< 30°)”, which is why a larger value is passed on with a warning rather than silently. And it is a worst case: “Depending on the signal waveform, the actual peak value deviation will normally be smaller than ΔP_max which is a worst-case estimate, combining the amplitudes and zero phase angles of two frequency components in the most unfavourable manner.”

Parameters

NameDescription
characteristic_phase_deviation_degOne or more characteristic phase deviations, in degrees, as characteristic_phase_deviation returns them.

Returns: The likely maximum peak-value deviation, in per cent. The printed ± is the sign of the deviation, not part of the size, so the number returned is the magnitude.

Raises

ExceptionWhen
ValueErrorIf a value is negative or not finite. Formula (6) prints the quantity inside absolute-value bars, so a negative one is not a characteristic phase deviation and is refused rather than folded.
UserWarningHumanVibrationWarning when a value exceeds the 30 degrees NOTE 2 limits the approximation to.
phase_tolerance_degrees(
name: str,
frequencies: ArrayLike,
) -> NDArray[np.float64]

The Table 5 characteristic phase-deviation band, region by region.

Footnote a of the table limits this to instruments that provide a measurement parameter not based on r.m.s. values, which is why it is a separate function rather than a third array beside the magnitudes: an r.m.s.-only meter is not graded on phase at all.

Parameters

NameDescription
nameOne of WEIGHTING_NAMES.
frequenciesFrequencies at which the band is wanted, in hertz.

Returns: The permitted deviation, in degrees, infinite in the two tails.

Raises

ExceptionWhen
ValueErrorAs for weighting_tolerance_percent.
PhaseVerification(
weighting: str,
frequencies_hz: NDArray[np.float64],
measured_phase_deg: NDArray[np.float64],
design_phase_deg: NDArray[np.float64],
deviation_deg: NDArray[np.float64],
characteristic_frequencies_hz: NDArray[np.float64],
characteristic_deviation_deg: NDArray[np.float64],
tolerance_deg: NDArray[np.float64],
within_tolerance: NDArray[np.bool_],
)

One measured phase response against the ISO 8041-1 Table 5 phase band.

Attributes

NameDescription
weightingThe weighting the response was measured for.
frequencies_hzThe frequencies it was measured at.
measured_phase_degThe measured phase, as supplied.
design_phase_degThe design-goal phase of Formula (H.1) at the same frequencies, on the branch Tables B.1 to B.9 print.
deviation_degThe phase error, measured minus design, in degrees.
characteristic_frequencies_hzThe frequencies the characteristic phase deviations are attributed to, which are all but the highest (H.2.1, Formula (H.3)).
characteristic_deviation_degThe characteristic phase deviation of Formula (6) at each of those, in degrees.
tolerance_degThe Table 5 limit at each of those, in degrees, infinite in the two tails.
within_toleranceWhether each characteristic phase deviation is inside its limit.

property

The frequencies whose characteristic phase deviation is too large.

property

Whether every characteristic phase deviation sits inside Table 5.

True says the phase response meets the criterion of footnote a of Table 5, which is the one an instrument reporting peak, MTVV or VDV is held to. An r.m.s.-only instrument is not graded on phase at all, and this is not a verdict on its magnitude response.

property

What this phase response costs a peak reading (Formula (H.4)).

The worst case over the measured range, which is where the maximum in the printed formula is taken.

PhaseVerification.plot(
ax: Axes | None = None,
*,
language: str = 'en',
**kwargs: Any,
) -> Axes

Draw the characteristic phase deviation inside the Table 5 band.

Requires matplotlib (pip install phonometry[plot]); returns the Axes.

Parameters

NameDescription
axExisting axes, or None to create a figure.
languageLabel language, "en" (default) or "es".
kwargsForwarded to phonometry._plot.vibration.plot_phase_verification.

Constant (mapping).

PVEM_INDICATION_TOLERANCES_PERCENT = {'indication': 4.0, 'low-frequency indication': 5.0, 'weighting consistency': 3.0}

Constant (mapping).

PVEM_MAX_EXPANDED_UNCERTAINTY_PERCENT = {'12.7': 2.0, '12.10.1': 2.0, '12.10.2': 3.0, '12.10.2 additional ranges': 4.0, '12.11.2': 4.5, '12.11.3': 3.0, '12.11.4': 5.0, '12.13': 3.0, '12.14': 2.0, '12.18': 0.01, '13.9': 5.0}

Constant (mapping).

REFERENCE_ACCELERATION_M_S2 = {'Wb': 1.0, 'Wc': 1.0, 'Wd': 1.0, 'We': 1.0, 'Wf': 0.1, 'Wh': 10.0, 'Wj': 1.0, 'Wk': 1.0, 'Wm': 1.0}

Constant (mapping).

REFERENCE_FREQUENCY_HZ = {'Wb': 15.915494309189533, 'Wc': 15.915494309189533, 'Wd': 15.915494309189533, 'We': 15.915494309189533, 'Wf': 0.3978873577297384, 'Wh': 79.57747154594767, 'Wj': 15.915494309189533, 'Wk': 15.915494309189533, 'Wm': 15.915494309189533}
reference_indication(name: str) -> float

The weighted indication a conforming meter shows at the reference.

Table 1 pairs each application with a reference frequency and a reference r.m.s. acceleration, and the weightings are not unity there, so the expected indication is the product of the two.

Parameters

NameDescription
nameOne of WEIGHTING_NAMES.

Returns: The weighted acceleration, in metres per second squared.

Raises

ExceptionWhen
ValueErrorIf the weighting is not one of the nine.

Constant (float).

RUNNING_RMS_CONSISTENCY_TOLERANCE_PERCENT = 2.0

Constant (tuple).

RUNNING_RMS_DECAY_RATE_DB_PER_S = ((0.125, 31.0, 40.0), (1.0, 3.8, 4.9), (8.0, 0.48, 0.62))
running_rms_decay_time(integration_time_s: float, *, method: str) -> float

When the running r.m.s. falls to 10 % after the signal is shut off.

5.13 (folio 20) applies a steady sinusoid at the reference frequency for at least 5 time constants (linear averaging) or 20 (exponential), shuts it off, and times the decay “from the start of the decay to the time at which the indicated value is less than 10 % of the initial value”.

Both averages have a closed form. The linear average of Eq. (2) keeps the last tau seconds of the record, so a time t after the cut its window still holds (tau - t) / tau of the original mean square and the indicated value falls as , reaching 10 % at t = 0,99 tau. The exponential average of Eq. (3) decays in power as , so the indication falls as and reaches 10 % at .

Both land inside the printed bands of RUNNING_RMS_DECAY_TIME_S for the three time constants the standard tabulates.

Parameters

NameDescription
integration_time_sThe averaging time tau, in seconds (> 0).
method"linear" (Table 10) or "exponential" (Table 11), the two averages of running_rms.

Returns: The time to 10 % of the initial indicated value, in seconds.

Raises

ExceptionWhen
ValueErrorIf method is neither average, or the integration time is not positive and finite.

Constant (mapping).

RUNNING_RMS_DECAY_TIME_S = {'linear': ((0.125, 0.124, 0.005), (1.0, 0.99, 0.05), (8.0, 7.92, 0.2)), 'exponential': ((0.125, 0.58, 0.03), (1.0, 4.61, 0.25), (8.0, 36.8, 2.0))}
RunningRmsDecayVerification(
*,
method: str,
integration_time_s: float,
measured_time_s: float,
printed_time_s: float,
tolerance_s: float,
)

One measured decay time against its row of ISO 8041-1 Table 10 or 11.

The row is the whole criterion: a printed time to 10 % of the initial indicated value and the tolerance printed beside it, for one averaging and one time constant. The verdict is derived from those fields rather than stored beside them, so a result cannot say it passed over numbers that do not.

Attributes

NameDescription
method"linear" (Table 10) or "exponential" (Table 11).
integration_time_sThe printed time constant the row is for, in seconds.
measured_time_sThe measured time to 10 % of the initial indicated value, as supplied, in seconds.
printed_time_sThe decay time the row prints, in seconds.
tolerance_sThe tolerance printed beside it, in seconds.

property

The measured time minus the printed one, in seconds.

property

The shortest decay time the row accepts, in seconds.

property

Whether the measured time sits inside the printed interval.

True says the time weighting decays as clause 5.13 requires, and nothing more: the other clauses a meter is graded on are hardware measurements this cannot stand in for.

RunningRmsDecayVerification.plot(
ax: Axes | None = None,
*,
language: str = 'en',
**kwargs: Any,
) -> Axes

Draw the measured decay time against the printed interval.

Requires matplotlib (pip install phonometry[plot]); returns the Axes.

Parameters

NameDescription
axExisting axes, or None to create a figure.
languageLabel language, "en" (default) or "es".
kwargsForwarded to phonometry._plot.vibration.plot_running_rms_decay_verification.

property

The longest decay time the row accepts, in seconds.

Constant (tuple).

SKIRT_TOLERANCE_PERCENT = (26.0, -21.0, 12.0)

Constant (tuple).

TAIL_TOLERANCE_PERCENT = (26.0, -100.0, inf)

Constant (mapping).

TRANSITION_FREQUENCIES_HZ = {'Wb': (0.251188643150958, 0.6309573444801932, 63.09573444801933, 158.48931924611142), 'Wc': (0.251188643150958, 0.6309573444801932, 63.09573444801933, 158.48931924611142), 'Wd': (0.251188643150958, 0.6309573444801932, 63.09573444801933, 158.48931924611142), 'We': (0.251188643150958, 0.6309573444801932, 63.09573444801933, 158.48931924611142), 'Wf': (0.05011872336272722, 0.12589254117941673, 0.3981071705534972, 1.0), 'Wh': (3.9810717055349722, 10.0, 794.3282347242813, 1995.2623149688789), 'Wj': (0.251188643150958, 0.6309573444801932, 63.09573444801933, 158.48931924611142), 'Wk': (0.251188643150958, 0.6309573444801932, 63.09573444801933, 158.48931924611142), 'Wm': (0.5011872336272722, 1.2589254117941673, 63.09573444801933, 158.48931924611142)}

Constant (float).

UNCONSTRAINED_BELOW = -100.0
verify_phase_response(
name: str,
frequencies_hz: ArrayLike,
measured_phase_deg: ArrayLike,
) -> PhaseVerification

Verify a measured phase response against the ISO 8041-1 Table 5 band.

Table 5 (folio 15) prints a characteristic phase deviation limit beside every magnitude limit: ±6° in the central region, ±12° in the two skirts and ±∞ in the two tails. The quantity those grade is not the phase error but Formula (6) of it, which is why this is a separate entry point from verify_weighting rather than a third array inside it: footnote a of the table applies the phase criterion only to instruments “that provide measurement parameters that are not based on r.m.s. values”, so an r.m.s.-only meter is never handed this verdict.

The design goal comes from Formula (H.1), which is the argument of the same H(s) frequency_weighting evaluates, and it is put on the branch Tables B.1 to B.9 print, so the measurement has to arrive on a continuous branch too. That is not a convenience: the two invariances the standard prints for this criterion hold on the continuous phase error and on no other. A constant phase error of c degrees is graded as c; a constant group delay, which is a phase error proportional to frequency, is graded as zero, and H.2.3.4 n) is explicit that “any remaining constant delay time (except 180°) does not influence the result at all”. Fold the error into a half turn either side of zero and the second one stops being true, which is why nothing is folded here. H.2.3.4 g) to m) is the standard’s own reconstruction of that continuous curve from wrapped phase-meter readings, and it belongs before this call: on a one-third-octave grid a wrap and a delay ramp are the same jump, so no check here could tell a measurement still carrying its wraps from the delay the criterion is meant to ignore.

What a polarity error looks like here, and why it is not this test. H.2.3.4 k) (folio 99) says a half-turn shift of every component “leaves the wave form of the signal unchanged, but would create catastrophic results attempting to apply the characteristic phase deviation (CPD) criterion”, and that the criterion “is not applicable to signal inversion. Signal inversion is a unique form of signal processing which needs its own test procedure, the polarity test”. An inverted measurement therefore fails this check with a 180-degree deviation everywhere, and is warned about, because the verdict is real but the diagnosis is a polarity test this function does not perform.

Parameters

NameDescription
nameOne of WEIGHTING_NAMES.
frequencies_hzThe frequencies the phase was measured at, in hertz, strictly ascending and at least two of them.
measured_phase_degThe measured phase at each frequency, in degrees, on the continuous branch Tables B.1 to B.9 print.

Returns: The verdict, as a PhaseVerification.

Raises

ExceptionWhen
ValueErrorIf the weighting is not one of the nine, if the two arrays do not have the same shape, if a frequency is not positive and finite, if a phase is not finite, or if there are fewer than two frequencies or they do not strictly ascend.
UserWarningHumanVibrationWarning when every phase error is a half turn, which is an inverted signal rather than a phase response the criterion can grade.
verify_running_rms_decay(
measured_time_s: float,
*,
integration_time_s: float,
method: str,
) -> RunningRmsDecayVerification

Verify a measured decay time against Table 10 or Table 11.

The verdict is one printed row: the measured time to 10 % of the initial value has to sit inside the printed interval for that time constant and that average. The rate column of Table 11 is deliberately not the criterion, for the reason RUNNING_RMS_DECAY_RATE_DB_PER_S explains.

Only the three time constants the two tables print can be checked, so an instrument averaging over any other time is refused rather than judged against a band the standard does not give.

Parameters

NameDescription
measured_time_sThe measured time to 10 % of the initial indicated value, in seconds (> 0).
integration_time_sThe averaging time it was measured at, which has to be one of the printed 0,125 s, 1 s and 8 s. Keyword-only, and so is the method: two times in seconds side by side are the kind of pair a positional call gets the wrong way round in silence.
method"linear" (Table 10) or "exponential" (Table 11).

Returns: The verdict, as a RunningRmsDecayVerification that keeps the printed row it was judged against.

Raises

ExceptionWhen
ValueErrorIf method is neither average, if the measured time is not positive and finite, or if the integration time is not one of the three printed time constants.
verify_weighting(
name: str,
frequencies: ArrayLike,
measured_factors: ArrayLike,
*,
expanded_uncertainty_percent: float | None = None,
) -> WeightingVerification

Verify a measured weighting response against ISO 8041-1 Tables 4 and 5.

The acceptance test is the one Annex B is written in: the deviation (measured / design - 1) * 100 at each frequency has to sit between the lower and upper limits of the region that frequency falls in.

Where the laboratory’s uncertainty goes. 13.1 (folio 42) and 14.1 (folio 48) print the same sentence: compliance is demonstrated when the result of a measurement of a deviation from a design goal, “extended by the actual expanded uncertainty of measurement of the testing laboratory”, does not exceed the specified tolerance limits, the uncertainty being calculated with the coverage factor k = 2 (ISO8041_COVERAGE_FACTOR). So the comparison is deviation + U <= upper and deviation - U >= lower: the band stays where Table 5 prints it and the measurement is what widens. That is not in conflict with 5.6.6 (folio 14), which says the Table 5 limits already include the applicable maximum permitted expanded uncertainties: the band is not widened by either sentence, and what 13.1 adds is the laboratory’s actual uncertainty, which is its own number and is bounded by MAX_EXPANDED_UNCERTAINTY_PERCENT.

Parameters

NameDescription
nameOne of WEIGHTING_NAMES.
frequenciesThe frequencies the response was measured at, in hertz.
measured_factorsThe measured weighting factors, linear and not in decibels, one per frequency.
expanded_uncertainty_percentThe testing laboratory’s actual expanded uncertainty of the deviation measurement, in per cent and already expanded with k = 2. None (the default) compares the bare deviation, which is the right reading only for a measurement whose uncertainty has been shown to be negligible.

Returns: The verdict, as a WeightingVerification.

Raises

ExceptionWhen
ValueErrorIf the weighting is not one of the nine, if the two arrays do not have the same shape, if a frequency is not positive and finite, if a measured factor is negative or not finite, or if the expanded uncertainty is negative or not finite.

Constant (float).

WEIGHTING_CONSISTENCY_TOLERANCE_PERCENT = 3.0
weighting_tolerance_percent(
name: str,
frequencies: ArrayLike,
) -> tuple[NDArray[np.float64], NDArray[np.float64]]

The Table 5 band around a weighting, region by region.

Parameters

NameDescription
nameOne of WEIGHTING_NAMES.
frequenciesFrequencies at which the band is wanted, in hertz.

Returns: (upper, lower) percentages, elementwise. lower is UNCONSTRAINED_BELOW outside the two outer transition frequencies, where the standard sets no lower limit at all.

Raises

ExceptionWhen
ValueErrorIf the weighting is not one of the nine, or a frequency is not positive and finite.
WeightingVerification(
weighting: str,
frequencies_hz: NDArray[np.float64],
measured: NDArray[np.float64],
design: NDArray[np.float64],
deviation_percent: NDArray[np.float64],
within_tolerance: NDArray[np.bool_],
expanded_uncertainty_percent: float = 0.0,
)

One measured weighting response against its ISO 8041-1 tolerances.

Attributes

NameDescription
weightingThe weighting the response was measured for.
frequencies_hzThe frequencies it was measured at.
measuredThe measured weighting factors, as supplied.
designThe design-goal factors of ISO 8041-1 Table 3 at the same frequencies.
deviation_percent(measured / design - 1) * 100 elementwise, which is the quantity the standard’s acceptance test is written in.
within_toleranceWhether each frequency is inside its band, with the deviation extended by expanded_uncertainty_percent as 13.1 and 14.1 require.
expanded_uncertainty_percentThe testing laboratory’s own expanded uncertainty, in per cent, that the verdict was reached with. 0,0 when the caller supplied none, which compares the bare deviation.

WeightingVerification.failing_frequencies_hz

Section titled “WeightingVerification.failing_frequencies_hz”

property

The frequencies whose deviation falls outside the band.

property

Whether every measured frequency sits inside its band.

True says the frequency weighting meets ISO 8041-1, and nothing more: the indication, linearity, overload and environmental tests of the same standard are hardware measurements this cannot stand in for.

WeightingVerification.plot(
ax: Axes | None = None,
*,
language: str = 'en',
**kwargs: Any,
) -> Axes

Draw the measured response inside the tolerance band.

Requires matplotlib (pip install phonometry[plot]); returns the Axes.

Parameters

NameDescription
axExisting axes, or None to create a figure.
languageLabel language, "en" (default) or "es".
kwargsForwarded to phonometry._plot.vibration.plot_weighting_verification.

WeightingVerification.worst_deviation_percent

Section titled “WeightingVerification.worst_deviation_percent”

property

The largest deviation in magnitude, signed as it was measured.