# Build a sound level meter

A sound level meter is not one algorithm but a short pipeline of them, and
IEC 61672-1 specifies every stage. phonometry implements each stage as an
independent, composable function; this page assembles them, in order, into a
working meter. Every snippet runs as written (the signals are synthesized so
the page is self-contained), and each stage links to the deep guide that
explains it fully.

<picture><source media="(prefers-color-scheme: dark)" srcset="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/diagram_slm_pipeline_dark.svg"><img src="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/diagram_slm_pipeline.svg" alt="The sound level meter pipeline of this page: the calibrator tone feeding sensitivity() to obtain the factor in pascals per digital unit, the measurement recording joining it as the calibrated pressure, and three readout branches from there, weighting_filter with time_weighting for the displayed LAF(t) and the percentile levels, laeq, sel and lc_peak for the integrated levels, and octave_filter for the one-third-octave spectrum, all closed by verify_weighting_class and verify_filter_class against the IEC 61672-1 Table 3 and IEC 61260-1 Table 1 acceptance limits" width="92%"></picture>

This is the same chain IEC 61672-1 draws for the physical instrument: the
class 1 calibrator anchors the microphone to 94 dB at 1 kHz, and every stage
that follows is one function of this page.

<picture><source media="(prefers-color-scheme: dark)" srcset="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/diagram_slm_chain_dark.svg"><img src="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/diagram_slm_chain.svg" alt="IEC 61672-1 sound level meter chain: a class 1 sound calibrator at 94 dB and 1 kHz coupled onto the measurement microphone with its windscreen, followed by the four instrument stages of microphone plus preamplifier, A, C or Z frequency weighting, squaring with the exponential F or S time weighting of 125 ms or 1 s, and the display of LAF(t) in dB re 20 µPa" width="92%"></picture>

The snippets on this page build on each other: run them top to bottom in one
session (or paste the whole page into a script).

## 1. The scenario

A meter needs two recordings from the *same* input chain: the calibrator tone
that anchors the digital numbers to pascals, and the measurement itself. Here
both are synthesized so you can run the page anywhere; in a real measurement
they come from your microphone.

```python
import numpy as np
from phonometry import filters, metrology, signals

fs = 48000

# Calibrator tone: 94 dB SPL = 1 Pa RMS at 1 kHz (IEC 60942).
#   Synthesized here; in the field, record a few seconds of your calibrator.
calibrator = np.sqrt(2) * np.sin(2 * np.pi * 1000 * np.arange(3 * fs) / fs)

# "Street" measurement: 10 s of pink background noise plus a 1 s horn-like
#   1 kHz event, so the statistical levels have something to separate.
recording = signals.noise_signal(fs, 10.0, color="pink", rms=0.02, seed=7)
recording[4 * fs : 5 * fs] += 0.2 * np.sqrt(2) * np.sin(
    2 * np.pi * 1000 * np.arange(fs) / fs
)
```

## 2. Calibrate: give the samples physical meaning

Digital samples are dimensionless; the **sensitivity factor** converts them
to pascals. `sensitivity()` computes it from the calibrator recording and, at
the same time, validates the recording's short-term stability the way
IEC 60942 qualifies the calibrator itself, so a badly coupled microphone is
caught here instead of corrupting every level downstream.

```python
cal = metrology.sensitivity(calibrator, target_spl=94.0, fs=fs)
# cal is in Pa per digital unit; every level function accepts it as
# calibration_factor. For this synthetic tone it is ~1.0.
```

Deep guide: [Calibration and dBFS](https://jmrplens.github.io/phonometry/signals/metrology/calibration/), which
also covers calibrating from a known microphone sensitivity and the digital
dBFS mode used when no physical reference exists.

## 3. Weight: frequency and time (IEC 61672-1)

The meter never shows raw pressure. The signal first passes the **A
frequency weighting** (the ear-response curve of IEC 61672-1), is squared,
and is then smoothed by the **Fast exponential detector** (time constant
125 ms). The result is the moving level a meter's display follows,
$L_{AF}(t)$:

```python
pressure = cal * recording                                # digital units -> Pa
weighted = filters.weighting_filter(pressure, fs, curve="A")
envelope = filters.time_weighting(weighted, fs, mode="fast")  # mean-square Pa^2
laf_t = 10 * np.log10(np.maximum(envelope, 1e-12) / (2e-5) ** 2)
# laf_t peaks near 80 dB during the event and settles near 55 dB between.
```

<picture><source media="(prefers-color-scheme: dark)" srcset="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/anim_time_weighting_dark.gif"><img src="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/anim_time_weighting.gif" alt="Animation: a tone burst driving the RC exponential detector, the capacitor charging and draining, while the Fast, Slow and Impulse meter needles follow their own ballistics" width="640" height="360" loading="lazy"></picture>

[Watch the high-resolution video (WebM)](https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/anim_time_weighting.webm)

The needle in the clip is that `time_weighting` call: a first-order low-pass
charging and draining on the squared signal. The three needles differ in one
number, the time constant, which is why Fast catches an event that Slow
smooths away. You rarely write this chain yourself: every level function of
the next step applies the frequency weighting internally, and the percentile
levels rebuild this Fast envelope for you. The energy metrics ($L_{eq}$, SEL)
integrate the squared weighted signal directly, with no ballistics at all —
which is why they show no needle movement to follow. The chain is shown here
because it *is* the meter's display.

Deep guides: [Frequency Weighting (A, C, Z)](https://jmrplens.github.io/phonometry/signals/levels/weighting/)
and [Time Weighting](https://jmrplens.github.io/phonometry/signals/levels/time-weighting/), which takes this same clip
apart against the IEC 61672-1 tone-burst table.

## 4. Integrate: the numbers a meter reports

One pass over the calibrated recording yields the standard readouts: the
energy-equivalent **$L_{Aeq}$**, the **percentile levels** that describe how
the level fluctuated ($L_{90}$ is the background, $L_{10}$ the events), the
**sound exposure level** that normalizes the event to one second, and the
C-weighted **peak** for impulsive content.

```python
la_eq = signals.laeq(recording, fs, calibration_factor=cal)     # ~70.2 dB
ln = signals.ln_levels(
    recording, fs, n=(10, 50, 90), weighting="A", calibration_factor=cal
)                                                # L10 ~78.0, L50 ~55.1, L90 ~54.9
lae = signals.sel(recording, fs, weighting="A", calibration_factor=cal)  # ~80.2
lc_pk = signals.lc_peak(recording, fs, calibration_factor=cal)           # ~84.4

print(f"LAeq {la_eq:.1f} dB | L10 {ln[10]:.1f} | L90 {ln[90]:.1f} "
      f"| LAE {lae:.1f} | LCpeak {lc_pk:.1f}")
```

Note the arithmetic the numbers encode: the 1 s event dominates $L_{Aeq}$ (it
sits 25 dB above the background, far more than the 10 dB the
nine-times-longer background gets back in duration), $L_{AE}$ is $L_{Aeq}$
plus $10\log_{10}$ of the 10 s duration, and $L_{90}$ barely notices the event at
all.

Deep guide: [Integrated and Statistical Levels](https://jmrplens.github.io/phonometry/signals/levels/levels/),
which adds noise dose and octave spectrograms; the $L_{den}$ and rating levels
continue in [Environmental levels](https://jmrplens.github.io/phonometry/environment/assessment/environmental-levels/).

## 5. Band-filter: the spectrum view (IEC 61260-1)

A class 1 meter with a filter set reports band levels. `octave_filter`
decomposes the calibrated signal into fractional-octave bands whose design is
anchored to the IEC 61260-1 band edges; `nominal=True` labels them with the
preferred frequencies you would read on an instrument.

```python
spl, bands = filters.octave_filter(
    recording, fs, fraction=3, nominal=True,
    calibration=filters.LevelCalibration(factor=cal),
)
# 33 one-third-octave band levels in dB SPL, labeled '12.5' ... '20k'.
# The '1k' band holds the event: ~70 dB, while its neighbors stay ~25 dB below.
print(dict(zip(bands, np.round(spl, 1))))
```

Deep guides: [Filter Banks](https://jmrplens.github.io/phonometry/signals/filters/filter-banks/) for the filter
architectures and zero-phase mode,
[Block Processing](https://jmrplens.github.io/phonometry/signals/filters/block-processing/) for streaming, and
[Multichannel and Performance](https://jmrplens.github.io/phonometry/signals/filters/multichannel/) for arrays.

## 6. Verify: is this meter class 1?

A real instrument is only a "class 1 sound level meter" after its weightings
and filters pass the acceptance limits of the standards. The library ships
the same verifiers it applies to itself in CI: `verify_weighting_class`
sweeps a `WeightingFilter` against the IEC 61672-1 Table 3 limits, and
`verify_filter_class` sweeps an `OctaveFilterBank` against the IEC 61260-1
Table 1 limits.

```python
wf = filters.WeightingFilter(fs, curve="A")
print(filters.verify_weighting_class(wf)["overall_class"])   # 1

bank = filters.OctaveFilterBank(fs, fraction=3)
print(filters.verify_filter_class(bank)["overall_class"])    # 1
```

The verdicts also come per band, so you can see exactly where a design would
leave its class corridor. Deep guides:
[Frequency Weighting](https://jmrplens.github.io/phonometry/signals/levels/weighting/) (section on class
verification) and [Filter class verification](https://jmrplens.github.io/phonometry/signals/filters/filter-compliance/) (the
Table 1 mask, class 0 and the compliance fiche).

## Where to go next

The meter built here is the trunk; the rest of the core grows from it.

- [Measurement uncertainty (GUM and Monte Carlo)](https://jmrplens.github.io/phonometry/signals/metrology/gum-uncertainty/):
  attach an uncertainty to the $L_{Aeq}$ you just computed, calibration term
  included.
- [Calibrated spectral analysis](https://jmrplens.github.io/phonometry/signals/spectra/spectral-analysis/): when
  bands are too coarse, the Welch PSD with confidence intervals.
- [Correlation, time delay and envelope](https://jmrplens.github.io/phonometry/signals/spectra/correlation-delay/):
  two microphones instead of one, and the delay between them.
- [Block Processing](https://jmrplens.github.io/phonometry/signals/filters/block-processing/): turn this page's
  offline meter into a streaming one with carried filter state.

## What this guide covers

**Covered.** This page composes stages implemented elsewhere into the pipeline
IEC 61672-1:2013 describes: the A frequency weighting and the Fast exponential
detector, $L_{Aeq}$, the percentile levels, the sound exposure level and the
C-weighted peak, the IEC 61260-1:2014 octave filters of `octave_filter`, and
the Table 3 (weighting) and Table 1 (filter) class acceptance limits checked by
`verify_weighting_class` and `verify_filter_class`. Each stage's own guide
states its coverage in detail.

**Not covered.** `verify_weighting_class` and `verify_filter_class` check the
frequency-response *design* of the digital filters against the standards'
tables. They do not run the IEC 61672-2:2013 pattern-evaluation tests a
physical instrument needs for type approval — self-generated noise, linearity
range, overload indication, directional response — nor the IEC 61672-3:2013
periodic tests a working instrument receives. A class verdict from this page
describes an algorithm, not a built device, and the screening checks of the
last section screen the *recording*, not the instrument. The IEC 60942:2017
calibrator conformance tests are not run here either; see
[Calibration and dBFS](https://jmrplens.github.io/phonometry/signals/metrology/calibration/) for exactly what
`sensitivity()` does and does not check.

## See also

- API reference: [`metrology.calibration`](https://jmrplens.github.io/phonometry/reference/api/metrology/calibration/),
  [`filters.weighting`](https://jmrplens.github.io/phonometry/reference/api/filters/weighting/),
  [`signals.levels`](https://jmrplens.github.io/phonometry/reference/api/signals/levels/),
  [`phonometry`](https://jmrplens.github.io/phonometry/reference/api/filters/phonometry/) and
  [`filters.compliance`](https://jmrplens.github.io/phonometry/reference/api/filters/compliance/).

## References

- International Electrotechnical Commission. (2013). *Electroacoustics —
  Sound level meters — Part 1: Specifications* (IEC 61672-1:2013).
  [IEC webstore](https://webstore.iec.ch/en/publication/5708).
  The blueprint of the instrument assembled on this page: the A frequency
  weighting and the Fast exponential time weighting of the level chain, the
  C-weighted peak and the sound exposure level, and the Table 3 class
  acceptance limits checked by `verify_weighting_class`.
- International Electrotechnical Commission. (2014). *Electroacoustics —
  Octave-band and fractional-octave-band filters — Part 1: Specifications*
  (IEC 61260-1:2014). [IEC webstore](https://webstore.iec.ch/en/publication/5063).
  The fractional-octave-band filters behind the spectrum stage, and the
  Table 1 class acceptance limits checked by `verify_filter_class`.
- International Electrotechnical Commission. (2017). *Electroacoustics —
  Sound calibrators* (IEC 60942:2017).
  [IEC webstore](https://webstore.iec.ch/en/publication/30045).
  The acoustic calibrator assumed by the sensitivity stage: the 94 dB
  principal level and the short-term stability check applied to the reference
  recording.