# Compliance and verification

Three kinds of evidence back a number computed with this library. The
**verifiers** are public functions that grade a filter or weighting you
configured against the acceptance limits of its governing standard. The
**conformance report** ([docs/CONFORMANCE.md](https://jmrplens.github.io/phonometry/reference/conformance/)) pins the
shipped implementation to the standards' own expected values and is
regenerated on every change. And the **scope notes** on each guide say where
the software's claim ends and a laboratory's begins. This page is the map of
the three.

## What a performance class asserts

IEC 61672-1:2013 (sound level meters) and IEC 61260-1:2014 (fractional-octave
filters) both specify two performance categories, **class 1 and class 2**,
and both define them the same way: the two classes share the *same design
goals* and differ mainly in the **acceptance limits** around those goals and
in the range of operational temperature; class 2 limits are greater than or
equal to class 1 limits everywhere (IEC 61672-1:2013 clause 1;
IEC 61260-1:2014 subclause 1.2).

A class is therefore not a quality grade but a **worst-case error bound
under stated conditions**, and the bound chains: a class 1 measurement needs
every stage at class 1, from the [calibrator](https://jmrplens.github.io/phonometry/signals/metrology/calibration/) through the
weighting to the band filter. One class 2 stage bounds the chain at class 2.

Two vocabulary edges: **class 0** belongs to the withdrawn IEC 61260:1995 /
ANSI S1.11-2004 masks, not to the 2014 edition (see
[Filter class verification](https://jmrplens.github.io/phonometry/signals/filters/filter-compliance/)); and a class
belongs to a specification, so the honest claim is "class 1 per
IEC 61672-1:2013 Table 3" or "class 1 per IEC 61260-1:2014 Table 1", never
"class 1" alone.

## The verifiers: which function proves what

One verifier per instrument stage, all sharing the same verdict vocabulary:
`overall_class` (the strictest class met, or `None`), per-band margins in
decibels to the nearest limit, and a `range_limited` flag when part of the
standard's frequency range could not be demonstrated at your sample rate.

| Stage | Verifier | Acceptance limits |
| :--- | :--- | :--- |
| Frequency weighting (A, C, Z) | `verify_weighting_class` | IEC 61672-1:2013 Table 3 |
| Weightings B and AU | `verify_weighting_class` | ANSI S1.4-1983 Tables IV/V; IEC 61012:1990 Table 1 |
| Fractional-octave filter bank | `verify_filter_class` | IEC 61260-1:2014 Table 1 (1995 mask via `edition="1995"`) |
| Intensity instrument spectrum | `verify_intensity_class` | IEC 61043:1993 Table 2 |

The masks are public too: `weighting_class_limits(1)` returns the Table 3
corridor and `class_limits(fraction, filter_class, omega)` the Table 1 one,
so a report can draw the limits a verdict was judged against. Verifying a
whole chain, in the configuration the conformance suite itself pins
(48 kHz, order 6, one-third-octave, 100 Hz to 10 kHz):

```python
from phonometry import filters

wf = filters.WeightingFilter(48000, "A")
weighting = filters.verify_weighting_class(wf)
print(weighting["overall_class"])       # 1
print(weighting["range_limited"])       # False

bank = filters.OctaveFilterBank(fs=48000, fraction=3, order=6,
                                limits=[100, 10000])
bands = filters.verify_filter_class(bank)
print(bands["overall_class"])           # 1
print(bands["range_limited"])           # True for a decimated bank
```

Read the flags, not just the class: `range_limited` is `True` on the
decimated bank because a band cannot be evaluated beyond its own processing
Nyquist, so the far stopband rests on the anti-alias argument rather than on
the band filter itself. The details, including why a +0.400 dB margin is the
ceiling for a passing Butterworth bank, are in
[Filter class verification](https://jmrplens.github.io/phonometry/signals/filters/filter-compliance/); the weighting
side is section 6 of [Frequency weighting](https://jmrplens.github.io/phonometry/signals/levels/weighting/).

<picture><source media="(prefers-color-scheme: dark)" srcset="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/weighting_class_mask_dark.svg"><img src="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/weighting_class_mask.svg" alt="A and C weighting deviations at 48 kHz threading within the IEC 61672-1 Table 3 class 1 acceptance corridor, with the wider class 2 limits dotted" width="80%"></picture>

When the verdict has to leave the console, `filter_class_compliance` wraps
the same verification as a result object with `.plot()` and `.report()`, the
one-page accredited fiche with an optional PASS/FAIL row against a required
class; `intensity_class_compliance` does the same for the IEC 61043
residual-index verdict:

```python
from phonometry import OctaveFilterBank, ReportMetadata, filter_class_compliance

bank_11 = OctaveFilterBank(fs=48000, fraction=1, order=6, limits=[125, 4000])
result = filter_class_compliance(bank_11)   # result.overall_class == 1
result.report(
    "iec61260.pdf",
    metadata=ReportMetadata(
        specimen="1/1-octave filter bank",
        measurement_standard="IEC 61260-1:2014",
        required_class=1,
    ),
)                                            # -> Class 1 - COMPLIES, PASS
```

## Reading the conformance report

The verifiers grade *your* configuration; the
[conformance report](https://jmrplens.github.io/phonometry/reference/conformance/) is the complementary evidence for
the implementation as shipped: each row pins one implemented quantity to the
governing standard's own expected value (a worked example or a
tolerance-table entry, not a regression baseline), with the computed value,
the signed delta and a pass/fail verdict. It is regenerated on every pull
request and CI fails if it drifts from the code.

Use it in three ways: find the row for a metric before trusting it (the row
names the standard, edition and clause the implementation was held to); note
that the class rows walk one pinned configuration, so section 2's verifiers
are how the claim transfers to your own bank; and cite the report of the
exact library version you pinned. Defects the re-derivation exposed in the
published standards themselves are in the [errata registry](https://jmrplens.github.io/phonometry/reference/errata/).

## What only a laboratory can attest

Both instrument series continue past their Part 1 with two test regimes, and
neither is something a software library can run on itself.

**Pattern evaluation** (IEC 61672-2:2013 for meters, IEC 61260-2:2016 for
filters) is the type-approval regime a *model* passes once: the tests
necessary to verify conformance to **all mandatory specifications** of the
Part 1 (IEC 61672-2:2013 clause 1; IEC 61260-2:2016 subclause 1.1), on
physical specimens — IEC 61260-2 requires at least three of them
(subclause 4.1). That includes everything a transfer function does not have:
static pressure, air temperature and humidity influence, electrostatic and
radio-frequency immunity, directional response, acoustical tests of the
weightings with the microphone in the sound field, level linearity of real
electronics, self-generated noise, tonebursts and overload behaviour
(IEC 61672-2:2013 clauses 7 to 9; IEC 61260-2:2016 clauses 7 to 9).

**Periodic tests** (IEC 61672-3:2013, IEC 61260-3:2016) are what a *working
instrument* receives in the laboratory, typically every year or two, on a
limited set of key tests "restricted to the minimum considered necessary",
valid for the environmental conditions of the day (IEC 61672-3:2013
clause 1; IEC 61260-3:2016 subclauses 1.2 and 1.3). Both Part 3 scopes end with the
same caveat: passing every periodic test supports **no general conclusion of
conformance** to the Part 1 unless the model's pattern approval is on record
(IEC 61672-3:2013 clause 1; IEC 61260-3:2016 subclause 1.5).

The same honesty applies to this library, in both directions. A verifier
verdict is a statement about a *design*: the digital transfer function you
configured fits the Part 1 acceptance mask, with the reported margins, over
the checked range; every measurement made wholly in software inherits it. It
is *not* a pattern evaluation, a periodic test, or a certificate for any
physical device: nothing here has a microphone, a temperature or a serial
number, and a real front end brings its own paper — the meter's periodic
test per IEC 61672-3 and the calibrator's conformance per IEC 60942, both
discussed in [Calibration and dBFS](https://jmrplens.github.io/phonometry/signals/metrology/calibration/). A defensible report
names both verdicts: the library's design verdict with the version and its
conformance report, and the instrument's test record with its date.

## Standards

IEC 61672-1:2013, *Sound level meters — Part 1: Specifications*: the two
performance categories and the Table 3 acceptance limits checked by
`verify_weighting_class`.
IEC 61672-2:2013 (*Pattern evaluation tests*) and IEC 61672-3:2013
(*Periodic tests*): the instrument test regimes delimited above, not run.
IEC 61260-1:2014, *Octave-band and fractional-octave-band filters — Part 1:
Specifications*: the Table 1 acceptance limits checked by
`verify_filter_class`.
IEC 61260-2:2016 (*Pattern-evaluation tests*) and IEC 61260-3:2016
(*Periodic tests*): the filter-set test regimes delimited above, not run.
IEC 61043:1993, *Instruments for the measurement of sound intensity*: the
class limits behind `verify_intensity_class`.

**Not covered.** The verification mathematics of each stage belongs to that
stage's own page; and every test of the Parts 2 and 3 themselves is
delimited here but not performed: no environmental, immunity, acoustical or
linearity test is run, and no physical instrument is assigned a class.